Determination of nitrite on manganese dioxide doped reduced graphene oxide modified glassy carbon by differential pulse voltammetry


Yılmaz-Alhan B., Çelik G., Oguzhan Caglayan M., ŞAHİN S., Üstündağ Z.

Chemical Papers, cilt.76, sa.8, ss.4919-4925, 2022 (SCI-Expanded) identifier

  • Yayın Türü: Makale / Tam Makale
  • Cilt numarası: 76 Sayı: 8
  • Basım Tarihi: 2022
  • Doi Numarası: 10.1007/s11696-022-02218-9
  • Dergi Adı: Chemical Papers
  • Derginin Tarandığı İndeksler: Science Citation Index Expanded (SCI-EXPANDED), Scopus, Chemical Abstracts Core
  • Sayfa Sayıları: ss.4919-4925
  • Anahtar Kelimeler: Differential pulse voltammetry, MnO2 doped reduced graphene oxide, Nitrite analysis, Sensors
  • Bilecik Şeyh Edebali Üniversitesi Adresli: Evet

Özet

© 2022, Institute of Chemistry, Slovak Academy of Sciences.A transition metal oxide, MnO2, doped reduced graphene oxide (rGO) was prepared to develop a 2D-carbon-metal oxide composite modified glassy carbon electrode (GC) for sensitive electrochemical determination of nitrite ion (NO2−). GO was first synthesized and reduced to rGO followed by a MnO2 doping process. The obtained MnO2-rGO composite was then characterized with Raman spectroscopy, infrared spectroscopy, and scanning electron microscopy techniques. The MnO2-rGO coated on GC electrodes were electrochemically characterized using cyclic voltammetry, electrochemical impedance spectroscopy and optimized for the determination of NO2− using differential pulse voltammetry. A linear calibration curve was obtained between 0.1 and 5.5 μM with LOD and LOQ values of 0.02 μM and 0.06 μM, respectively. The selectivity of the proposed sensor was tested in different substances and no significant interference was found. The sensor validation test showed that the precision (% RSD) and accuracy of the system were around 1.95–2.73% and (− 3.5)–2.5%, respectively. Finally, real sample tests with commercially available juice samples and tap water confirmed that the method could detect the spiked NO2− in real samples. As a result, the sensitive and easy determination of NO2− has been achieved using MnO2-rGO composite materials in real samples with good recovery values and minimum interference.